13. Bump Testing & Calibrating The ToxiRAE Pro PID
Note: Calibration intervals and bump test procedures may vary due to national
legislatio n.
13.1. Bump Testing & Calibrating With An AutoRAE 2™
Bump testing and calibrating a ToxiRAE Pro PID can be done with an AutoRAE 2
Automatic Test And Calibration Station. Follow all instructio ns in the AutoRAE 2 User’s
Guide for successful bump testing or calibration.
13.1.1. Bump (Functional) Testing
With the ToxiRAE Pro PID in Normal Mode:
1. Connect the calibration gas cylinder, flow regulator, and calibration adapter to the
ToxiRAE Pro PID and start the gas flow.
2. Make sure the unit goes into alarm and that the buzzer produces at least two beeps per
second, the LED lights flash on and off, and the vibration alarm functions. The
display backlight should illuminate and an alarm message should be shown in the
display.
3. Turn off the gas flow.
4. Remove the calibration adapter.
Note: Bump testing and calibration can also be performed by using the RAE Systems
AutoRAE 2 Automatic Test And Calibration Station. Refer to the AutoRAE 2 Users
Guide for instructions.
13.2. Calibration Adapter
The ToxiRAE Pro PID must be calibrated using a fixed-flow regulator with a flow rate
between 0.5 and 1.0 liters per minute. The ToxiRAE Pro PID is supplied with a special
calibration adapter that covers the gas fan inlet. If the local air has impurities that might
interfere with zero (fresh air) calibration, the calibration adapter and a cylinder of clean
zero air should be used.
13.2.1. Connecting The Calibration Adapter
1. Place the Calibration Adapter on the top cap of the ToxiRAE Pro PID by pressing
it into place.
2. Attach the hose from the regulator/calibration gas cylinder to the Calibration
Adapter’s inlet.