ToxiRAE Pro PID Users Guide
45
17. Exit ProRAE Studio II.
18. Press [Y/+] on the ToxiRAE Pro PID to exit Communication Mode.
12.2. Deactivating Policy Enforcement
12.2.1. AutoRAE 2 Cradle
To deactivate Policy Enforcement when using an AutoRAE 2 Cradle, follow the
procedure for changing settings. See page 39 for details.
12.2.2. ToxiRAE Pro PID Desktop Cradle Or Travel Charger
If the ToxiRAE Pro PID screen displays the message that it must be bump tested or
calibrated, and if the option to bypass bump testing or calibration is not available, you
should shut off the instrument and follow the procedure outlined here if you want to
change the Policy Enforcement settings:
1. Use a USB cable to connect the ToxiRAE Pro PID in its Travel Charger or Desktop
Cradle to a computer running ProRAE Studio II.
2. Enter Diagnostic Mode on the ToxiRAE Pro PID (with the instrument turned off,
press and hold [Y/+] and [MODE] until it starts up.
3. After startup, enter the password when prompted (default is “0000”) and press
[MODE].
4. Press [MODE] repeatedly until you see the “Enter Communications Mode?” screen.
5. Press [Y/+] to enter Communications Mode.
6. Start ProRAE Studio II.
7. Select “Administrator.
8. Input the password (the default is “rae”).
9. ClickOK.
10. Click A (detect instruments automatically).
11. Click on the instrument’s icon when it appears.
12. Click “Select.”
13. Click “Setup.”
14. Click Policy Enforcement. The Policy Enforcement pane is shown.
15. Deselect Policy Enforcement features you do not wish to use.
16. Click “Upload all settings to the instrument.