29
When the Man Down feature is on and there is no gas alarm, the ToxiRAE Pro PID senses that
it is motionless for the amount of time set in theMotionless Time parameter. If the
instrument is not moved during that time, then a pre-alarm is activated to alert the user, and
shows the “Are You OK?” screen. Pressing [Y/+] clears the alarm and returns the ToxiRAE
Pro PID to its normal operation. Pressing [MODE] sets it into Man Down Alarm (and if
wireless connectivity is enabled, a Man Down message is sent in real time to remote
observers). If neither key is pressed, then after the countdown, it goes into Man Down Alarm
(again sending a message to remote observers if wirelessly enabled).
Settings are available for:
Off/O n
Motionless Time (time the instrument is motionless before initiating a pre-alarm)
Sensitivity (set to low, medium, or high to compensate for ambient vibration or motion)
Warning Time (countdown, in seconds, from pre-alarm to Man Down alarm)
When the Man Down alarm is activated, the buzzer sounds and LEDs flash continuously, and a
countdown begins.
If the ToxiRAE Pro PID’s user presses [Y/+] for “Yes” in response to the Are You
OK?” question on the screen before the countdown reaches zero, the Man Down alarm
stops and the main reading screen is displayed.
If the person does not press [Y/+] for Yes” in response to the Are You OK?”
question on the screen before the countdown reaches zero, the Man Down alarm is
sounded and LEDs flash continuously.
If the person presses [MODE] during the countdown, answering the “Are You OK?
question with “No,the Man Down alarm starts.
If the ToxiRAE Pro PID’s wireless connectivity is enabled, a Man Down message is also sent to
remote observers.
Exit (the Alarm Settings Submenu)
Press [Y/+] to exit the Alarm Settings submenu. Alarm Settings is exited, and Datalog is
highlighted.
If you want to cycle through the Alarm Setting submenus, press [MODE] instead of [Y/+].