28
Buzzer & Light
The buzzer, light, and vibration alarms can be programmed so that any combination, all,
or none of them are activated when the ToxiRAE Pro PID goes into alarm.
Important! If you program the ToxiRAE Pro PID so that no alarms are activated in an
alarm conditio n, you may put yourself or your workers at risk of being exposed to a
hazardous atmosphere and being unaware of it, as the indicatio n of a dangerous condition
will be limited to the concentration shown in the display and the name of the type of
alarm flashing on the display.
1. Choose an alarm combination setting by pressing [MODE] until your choice is
highlighted.
2. Press [Y/+] to select it.
3. Press [MODE] repeatedly until “Exit” is highlighted.
4. Press [Y/+] to register your selection and exit.
To exit without saving the new value:
1. Make sure your previous setting is selected.
2. Press [MODE] repeatedly to advance to Exit.
3. Press [Y/+] to exit.
Man Down Alarm
The Man Down Alarm is a critical and potentially lifesaving safety feature of every
ToxiRAE Pro PID. The Man Down Alarm is based on the premise that if the instrument
is motionless when it is not supposed to be, something wrong may be happening to its
user. If that is the case, the ToxiRAE Pro PID not only goes into alarm locally on the
instrument, but also remotely, over the RAE Systems Dedicated Wireless Network, to
notify people in the vicinity, as well as remote safety officers at a command center, that a
person is down, so that help can be dispatched quickly.
Whenever the Man Down feature is on, the main screen displays a Man Down icon along the
top to indicate it is active: