User Manual

Table Of Contents
ToxiRAE Pro PID Users Guide
8
2. Specifications
ToxiRAE Pro PID Specifications
118 mm x 60 mm x 30 mm (4.6" x 2.4" x 1.2")
235g (8.3 oz)
Internal rechargeable Li-ion
Less than 4 hours
More than 12 hours on full charge (normal operation)
Charging and PC communication cradle: AC / DC adapter with international
prong kit (100 to 240V, 50/60 Hz AC to 12V DC)
Multi-unit charging station for up to 5 units (data download from one):
AC/DC adapter for the multi-unit charging station
Audio alarm 95 dB @ 30cm
Visual alarm 4 bright red LED lights
Vibration alarm
Activated display backlighting and on-screen indication of alarm conditions
Wireless alarm notification sent to remote observers (wireless versions only)
Monochrome dot-matrix graphical LCD display
Manually and on alarm
2 programming/operation keys
Real-time reading of VOC concentrations in parts per million or mg/m
3
;
battery status; data logging on/off status; wireless reception quality; STEL,
TWA, and peak values; calibration and measurement gases, correction factors
Fan-assisted diffusion
Capacity: 3 months at 1-minute interval
Datalog interval: user-configurable from 1 to 3,600 seconds
Two-point field calibration of zero and standard reference gas
Adjustable span value
Maintains calibration date
IP-54
AutoRAE 2 Automatic Test and Calibration Station
Password-protected calibration settings, alarm limits, and data
Hazardous
Location
Approvals
Class I, Div. 1; Gr. A,B,C,D T4
Class I, Zone 0 A/Exia IIC T4
ATEX:
II 1G,
Ex ia IIC Ga T4
IECEx: Ex ia IIC Ga T4
China: Ex ia IIC T4
(European
Conformity)
EMC directive: 2004/108/EC
R&TTE directive: 1999/5/EC
ATEX directive: 94/9/EC
-4° to 131° F (-20° to 55° C)
0 to 95% relative humidity (non-condensing)
Two years on non-consumable components (per RAE Systems Standard
Warranty); one year on fan, sensor, battery, and other consumable components