Operation Manual

ToxiRAE Pro User’s Guide
47
3. Press [MODE] if you want to quit without calibrating.
During calibration, the calibration gas name, its concentration, and the countdown are
displayed. Depending upon the device alarm settings, the monitor may go into low and
then high alarm once span calibration is complete, which is expected in response to the
test gas.
Note: You can abort span calibration during its countdown by pressing [MODE].
Upon completion, the screen “Span is done!” appears along with the reading (it should be
within 10% of the span gas concentration).
Important! After calibration is complete, you must remove the Calibration Adapter from
the ToxiRAE Pro. If the Calibration Adapter is attached during normal operation,
inconsistent and lower-than-normal readings will occur because of decreased
concentration of the gas being monitored.
14.4.1. Changing Span Gas Concentration
If you selected “Change” for span gas concentration, the following screen appears:
1. Press [Y/+] repeatedly to select a desired number. Numbers increase from 0 to 9.
Once 9 is reached, pressing [Y/+] again wraps” around back to 0.
2. Step from one digit to the next by pressing [MODE].
3. When you are done, press [MODE] until either “OKor Cancelis selected.
Press [Y/+] to register your selection (if OKis highlighted) or to exit without
registering it (if Cancel” is selected).