ToxiRAE Pro User’s Guide
35
Temp. Unit
The display unit of the internal temperature sensor can be switched between Fahrenheit
and Celsius.
To change the Temp. Unit:
1. Press [MODE] to step from one option to the other. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
the option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Temp. Unit menu, and advance to
Language.
Language
English is the default language, but other languages can also be selected for the
instrument.
1. Press [MODE] to step from one option to the next. The highlighted option
changes each time you press [MODE].
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
this option has been selected).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Language submenu, and advance to
Zero At Start.
Zero At Start
When Zero At Start is turned on, the instrument starts with zero calibration every time it
is turned on.
1. Press [MODE] to choose between “Off” and “On.”
2. Press [Y/+] to make your selection (the dark circle in the “radio button” indicates
your selection).
3. When you have completed your selection, press [MODE] repeatedly until “Exit”
is highlighted.
4. Press [Y/+] to save your selection, exit the Zero At Start submenu, and advance to
LCD Contrast.
LCD Contrast
The display’s contrast can be increased or decreased from its default setting. You may
not need to ever change the default setting, but sometimes you can optimize the display to
suit extreme temperature and ambient brightness/darkness conditions.
1. Press and hold [Y/+] to select a desired level of contrast. The range is 1 to 100.
Once 100 is reached, pressing [Y/+] again "wraps" around back to 1.