Problem Cause Solution
peak gas exposure reading.
Alarm setpoints are set
incorrectly.
Define the alarm setpoint in
Device Configurator.
Sensors require calibration. Carry out calibration.
Missing or faulty sensors. Replace the sensors.
Battery temperature is out
of acceptable range.
Move to lower temperature
ambient to charge the battery.
Battery indicator doesn’t
display when charging.
Battery is depleted.
Charge the battery for 8 hours.
If the battery indicator doesn’t
light after charging, contact
Honeywell
Battery icon flashes in
blue.
Battery is out of required
charging temperature
range.
Move to 0~45°C ambient
temperature.