Bump Test
The detector can be tested in two ways:
• Apply gas from a cylinder to the sensors manually through the calibration cap, and using the
Safety Suite Device Configurator (SSDC) software or the Device Configurator (DC) app.
• Use an IntelliDoX module. For further reference see the IntelliDoX User Manual.
CAUTION
Move to a normal atmosphere (20.9% v/v O
2
) that is free of hazardous gas. Use 50% LEL for
test gas.
Details for Bump Test and maintenance:
l Recommendations for initial checking of the equipment on a routine basis including the
maximum time interval between calibrations.
l Perform a functional check with gas before each day of use.
l Honeywell recommends bump testing the sensors before each day’s use to confirm their
ability to respond to gas by exposing the apparatus to a gas concentration that exceeds the
alarm setpoints. Manually verify that the audible and visual alarms activate.
l The combustible sensor is factory calibrated to 50% LEL methane. If monitoring a different
combustible gas in the % LEL range, calibrate the sensor using the appropriate gas.
Procedure to Bump Test the detector via the calibration cap and the DC app on
a mobile device
1. Turn On the BW Icon. Place the cap over the detector, and then press down on both tabs
to snap it into place. Wait a few minutes to sensors warm up.
BW Icon 19 User Manual