User's Manual

AreaRAE Plus & AreaRAE Pro User’s Guide
104
19.6.2 General Alarms
Message
Condition
Alarm Indications
HIGH
Gas exceeds High Alarm limit
3 beeps/flashes per second
OVR
Gas exceeds sensor’s measurement range
3 beeps/flashes per second
MAX
Gas exceeds electronic circuit’s maximum
range
3 beeps/flashes per second
LOW
Gas exceeds Low Alarm limit*
2 beeps/flashes per second
TWA
Gas exceeds TWA” limit
1 Beep/flash per second
STEL
Gas exceeds STEL” limit
1 Beep/flash per second
Crossed pump
icon flashes
Inlet blocked or pump failure
3 beeps/flashes per second
Lamp”
flashes
PID lamp failure
3 beeps/flashes per second
Empty battery
icon flashes
Low battery
1 flash, 1 beep per minute
CAL
Calibration failed, or needs calibration
1 beep/flash per second
NEG
True sensor reading is below zero, even
though a zero reading is shown for the
sensor.
1 beep/flash per second
* For oxygen, low alarm limit” means a concentration is lower than the low alarm limit.