When the Man Down alarm is activated, the buzzer sounds and LEDs flash continuously, and a countdown
begins.
If the MicroRAEs user presses [Y/+] for “Yes in response to theOK?” question on the
screen before the countdown reaches zero, the Man Down alarm stops and the main reading
screen is displayed.
If the person does not press [Y/+] for Yes” in response to the OK? question on the screen
before the countdown reaches zero, the Man Down alarm is sounded and LEDs flash
continuously.
If the person presses [MODE] during the countdown, answering the “OK? question by
pressing [MODE] for “No,the Man Down alarm starts.
If wireless connectivity is enabled, and the QRAE 3 is connected to a network, a Man Down message is
also sent to remote observers.
7.7. Calibration Status
If any sensor requires calibration, then “Calibration Overdue” icon is shown by the sensor name
in the display:
Calibration is required if:
The sensor module has been replaced with one whose calibration is overdue.
The defined period of time between calibrations has been exceeded, according to the
policy set for the instrument.
If you have changed the calibration gas type without recalibrating the instrument.
The sensor has failed a previous calibration.