User manual
Default form of the report from formulation template: 
---------------------------------------------- 
Recipe 
---------------------------------------------- 
{40:Start date:,-25}{180} 
{40:End date:,-25}{181} 
{40:Name:,-25}{175} 
{40:Code:,-25}{176} 
{40:Status:,-25}{182} 
{40:Measurements:,-25} 
---------------------------------------------- 
{185:(50,-20) (7)(11) 
(40:Nominal Mass:,-25)(186)(11) 
(40:Difference:,-25)(187)(11) 
---------------------------------------------- 
}{40:Mass:,-25}{184}{11} 
---------------------------------------------- 
The report from each completed formula making process is simultaneously 
saved in the database of <
Reports From Formulas>, where the files are 
named by their date and hour of process execution and formulation status. 
List of data for the formula making process – see ch. 31.7.6 of this user 
manual. 
28. WORKING MODES – DENSITY 
Working mode < Density> enables determining density of solids, liquids 
and materials of high viscosity. The density is determined on basis of 
Archimedes principle, according to which any floating object displaces its 
own weight of fluid. The density mode also enables utilizing a pycnometer 
for determining density of liquids. 
28.1. Starting the working mode 
Procedure: 
•  While in main window of the software, press 
 on-screen key, located 
in the upper bar of the window, which opens submenu <Working 
Modes> containing list of available working modes, 
•  Select working mode <
Density>, the software automatically returns 
to displaying the main window, and the upper bar indicates the icon of 
selected working mode, 
99