•  Enter parameter group <  Working modes> according to ch. 
23 of this manual, choose: „
 Counting Pieces /   Automatic 
correction of reference mass” and set appropriate option. 
Options: 
No 
-  Automatic correction of reference mass disabled 
Yes 
-  Automatic correction of reference mass enabled 
Function <
 Automatic correction of reference mass> in mode 
<
 Counting Pieces> is enabled at the moment of estimating the sample 
quantity and signalled by displaying <PCS> and <SMP> (single piece mass) 
on the top part of the display. 
There are four criteris of working “Automatic correction of reference 
mass”
 function: 
1.  equilibrium should be reached, 
2.  quantity of pieces should be increased, 
3.  added quantity of pieces should not be greater than double 
number of pieces on the pan, 
4.  the new sample can be different from the old sample by ± 0.3 
of pcs (absolute value), 
If a user recognises that the sample quantity is adequate the unit mass 
(single piece mass) can be saved (see ch. 24.5 of this manual) 
and disable the function by pressing ENTER/PRINT. 
Notice: 
While the function is active ENTER/PRINT changes its functionality. 
Pressing ENTER/PRINT does not result in printing and saving weighings. 
25.2.2. Minimum reference mass 
It concerns the 
<
 Counting pieces> working mode 
Users before beginning the procedure of single piese mass evaluation can 
declare “minimum reference mass” i.e. minimum total weight of all pieces 
put on the scale pan expressed in reading divisions. 