User manual
•  Simultaneously, the workspace of the display indicates a message: 
<Start estimating density>. 
28.2. Local setting of a working mode 
Local settings for the working mode <
Density> are accessible on  
pressing a hot key <
 Local parameters>: 
Standard liquid 
Function dedicated for determining the standard liquid. 
There are three options: water, alcohol and other. If option 
“other” is selected, then its density should be specified. In 
case of the other two types of liquids, their ensity values 
are implemented in the software 
Temperature 
Parameter of the standard liquid inserted by the scale 
user. Declared temperature of the standard liquid 
automatically determines its density based on a table of 
liquid densities. In case of the liquid type “Other”, the 
temperature parameter is not applicable, the density value 
is inserted manually 
Standard liquid 
density 
Parameter designed for manual determining the density of 
standard liquid in [g/cm
3
] 
Sinker volume 
Parameter implemented for manual determining the 
volume of a sinker expressed [cm
3
] 
Ask about sample 
number 
Function ordering inserting number of a sample before 
initaining the test procedure 
Pycnometer mass 
Parameter designed for manual determining mass of a 
pycnometer expressed in [g]. If inserted value is equal to 
„0” then on test procedure start, the pycnometer mass 
shall be measured. 
Pycnometer density 
Parameter intended for manual determining volume of the 
pycnometer expressed in [cm
3
] 
Unit 
Weighing unit set to read the measurement result. This 
weighing unit is used for presenting measurement result, 
data in reports, databases and summaries. 
Save mode 
Detailed description in ch. 23.1 of the user manual 
Checkweighing 
Detailed description in ch. 23.3 of the user manual 
Tare mode 
Detailed description in ch. 23.4 of the user manual 
Statistics 
Detailed description in ch. 23.6 of the user manual 
100