User manual
Return to weighing:
37
The changes introduced are saved for good after they are confirmed.
Press
several times until the
following message box appears:
Press:
– to confirm changes or – to abort
changes. The program returns to weighing.
17.2. Start mass adjustment
It is possible to adjust only a start mass, it helps to correct the start zero
when the span does not change.
Procedure:
• Enter submenu <
Customer Calibration> according to ch. 17
and select: “
Platform 1 / Setting of start mass”,
• After entering the parameter the following message box appears:
• Take the load off the pan of platform 1,
• Press button . The following message appears during adjusting
start mass: „Evaluation of starting mass”,