16.5. Software version
Procedure:
• Enter submenu < Others> according to ch. 16 of this manual,
choose <
Software Version> to see the program version
number.
16.6. Factory number
Procedure:
• Enter submenu < Others> according to ch. 16 of this manual,
choose <
Factory number> to see the program version number.
17. CUSTOMER CALIBRATION
An option only for non-verified scale
Scales require to recalculate internal divisions to more suitable ones (e.g. g,
kg etc.). In order to do this they require a calibration factor. It is adjusted
during the calibration procedure using a mass standard. Calibration should
be made when weighing a standard mass shows a different mass value.
To enter < Customer Calibration>, press and then:
< Customer Calibration>.
17.1. Calibration procedure
• Enter submenu <
Customer Calibration> according to ch. 17
and select: “
Platform 1 / Calibration”,
• After entering the parameter the following message box appears:
35