Caution:
Users can terminate testing anytime by pressing the screen button
(stop process) situated in the bottom part of the scale display.
During testing the rest of screen buttons, function keys,
and are not active.
Put the “A” standard mass on the pan,
After stabilisation (signalled by ) confirm the measurement by
pressing the
,
The process command changed from <Put A1> to <Put B1>,
Take off the "A" mass standard and put a "B" standard on the pan,
After stabilisation (signalled by ) confirm the measurement by
pressing the
,
The process command changed from <Put B1> to <Put B2>,
Take off the “B” mass standard and put it again,
After stabilisation (signalled by ) confirm the measurement by
pressing the
,
The process command changed from <Put B2> to <Put A2>,
Take off the “B” mass standard and put the “A” standard,
After stabilisation (signalled by ) confirm the measurement by
pressing the
,
Automatically some descriptions will change: weighed standard
to <A1> and number of measurements for the measurement
series to <P 2/1 0>.
Repeat measurements for the number of measurements declared
in the submenu <
Number of measurements>.
After the last measurement in a series is confirmed, the message
window is displayed: