4.1. Comparator WPY/KB – preparation for work 
Comparators of WPY/KB series, because of its construction (sensitive strain 
gauges), require appropriate conditions to work properly, including stable 
temperature. Temperature drifts in rooms with comparators should not 
exceed 1°C per hour. 
Proceeding in case of preparing the comparator to work:
1.  After unpacking and taking off transport protections put the scale 
in the destination place and level the komparator platform. 
2.  The room temperature must be stable – maximum allowable 
temperature change is 1°C / h – and protected from accidental draughts 
air flows e.g. from air conditioning directed on the comparator pan. 
3.  Then switch on the device and leave unloaded and on for 4 hours. This 
is warming time during which electronical and mechanical parts reach 
stable temperature conditions. 
4.  After 4 hours in stable ambient conditions the comparator can be 
prepared for adjustment. 
5.  First load the comparator pan to the maximum capacity 1000kg and 
keep this state at least 2.5 minuty (150 seconds). 
6.  Secondly unload the pan and wait at least 2 minuts (120 seconds), 
to allow the scale to return to the minimum stable value close to „zero”. 
7.  The operation of initial loading should be repeated 5 times, every time 
in reference to time intervals given above. 
8.  After the 5-time initial loading the comparator can undergo the 
adjustement procedure: 2 minutes after unloading the pan and start 
the adjustment procedure from factory menu – first start mass is 
determined then load the calibration weight following hints that appear 
on the display. Confirm the stable result after 2 minutes. This way the 
adjustement procedure will be completed. 
9.  The comparator is ready for work. 
Notice: 
If working conditions are changed (temperature change greater than 1°C, 
humidity change, sudden long lasting air flow causing a change of ambient 
conditions) the adjucement procedure described above needs to be repeated. 