User manual
Scale parameters - Weighing 
Icon Description  Value 
 Platform 1  - 
Median Filter  None 
Filter Slow 
Autozero No 
LO threshold  0.1 
Working modes - Comparator – Komparator 
Icon Description  Value 
 Comparator  - 
Save Mode  Automatic first stable 
Checkweighing No 
Tare mode  Single 
Statistics Global 
Method ABBA 
Number of measurements  6 
Ask for numbers of standards  No 
3.  Start the comparison process and follow directions appearing on the 
display of indicator. 
Notice: 
Before you put the first weight on the pan you need to zero the device 
manually by pressing button 
 on the indicator keypad. 
4.  Put on the pan the first standard weight “A”. On the display you will see 
“-----“ signalling the process of stabilization of the measurement system. 
81