User manual
a)  Estimate the unit mass (see 24.2 and 24.3), 
b)  Enter the products database 
, 
c)  Keep the finger pressed on the required position then a context 
menu is displayed, 
d) Choose option <Ascribe standard>, then the standard unit mass  
is attributed to the product in the field <Mass>. 
Notice: 
Attributing a standard to a selected product is also possible by 
programmable button. Setting programmable buttons is described in ch. 
17.2 of this manual The list of functions is described in APPENDIX B 
of this manual. 
25. WORKING MODES – DEVIATIONS 
The program allows to check weighings in deviations (in %) around an 
outlined standard mass. The standard mass can be outlined by weighing 
or entering it by a user. 
25.1. Starting the operating mode 
Procedure: 
•  While in the main window press   in the top bar of the window then 
you will see a submenu <Working Modes> comprising a list of modes, 
• Choose mode <
 Deviations>, the program will automatically 
returns to the main window displaying icon 
 in the top bar, 
•  The weight unit is automatically changed to „%” and two screen 
buttons on the right side appear: 
Enter reference mass 
Estimate reference mass 
77