User manual
20.3. Sound signal
Procedure:
• Enter < Others> according to ch. 20 of this manual, < Beep>
and set accordingly.
Settings:
None
- Sound for buttons and proximity sensors disabled
Buttons
- Sound for buttons enabled
Sensors
- Sound for proximity sensors enabled
All
- Sound for buttons and proximity sensors enabled
20.4. Touch panel calibration
Touch panel calibration is required when inappropriate operation is
recognized. E.g. the reaction in a different place than the touching point.
Procedure:
• Enter submenu < Others> according to ch. 20 of this manual,
• Select <
Touch Screen Calibration> and then an editing field
appears,
• Using a thin and soft pointer press (keep pressed for some time)
in the point where the cross appears, after indicating the 4
th
place
confirm changes by pressing
.
21. CUSTOMER CALIBRATION
An option only for non-verified scale
Scales require to recalculate internal divisions to more suitable ones (e.g.
g, kg etc.). In order to do this they require a calibration factor. It is adjusted
during the calibration procedure using a mass standard. Calibration should
be made when weighing a standard mass shows a different mass value.
59