User manual
108
Technical data:  WPY 500/KB 
Max capacity  510kg 
Readability 
2g (for mass standards 200kg) 
5g (for mass standards 500kg) 
Repeatability * 
2g (for mass standards 200kg) 
4g (for mass standards 500kg) 
Electric compensation range  0 ÷ 510 kg 
Supplementary weights  none 
Stabilization time  120 s 
Adjustment / calibration  External 
Power supply  110 ÷ 230V AC / 50 ÷ 60Hz / 13,5 ÷ 16V DC / 1,1A 
Pan size 
φ1000 mm 
Weighing unit dimensions  1347x1166x200 mm 
Control unit dimensions  206x140x70 mm 
Net weight  200 kg 
Ambient conditions: 
Working temperature  +15°C ÷ +35°C 
Change rate of working temperatures 
±0,5°C/h 
Atmospheric humidity  30% ÷ 70% 
Change rate of atmospheric humidity 
±10%/4h 
Technical data:  WPY 1000/KB 
Max capacity  1100kg 
Readability 
5g (for mass standards 500kg) 
10g (for mass standards 1000kg) 
Repeatability * 
4g (for mass standards 500kg) 
8g (for mass standards 1000kg) 
Electric compensation range  0 ÷ 1100 kg 
Supplementary weights  none 
Stabilization time  120 s 
Adjustment / calibration  External 
Power supply  110 ÷ 230V AC / 50 ÷ 60Hz / 13,5 ÷ 16V DC / 1,1A 
Pan size 
φ1000 mm 
Weighing unit dimensions  1347x1166x200 mm 
Control unit dimensions  206x140x70 mm 
Net weight  200 kg 
Ambient conditions: 
Working temperature  +15°C ÷ +35°C 
Change rate of working temperatures 
±0,5°C/h 
Atmospheric humidity  30% ÷ 70% 
Change rate of atmospheric humidity 
±10%/4h 
*   Repeatability is expressed as a standard deviation of six ABBA cycles 
(according to R111 OIM) in stable laboratory conditions. 










