User manual

108
Technical data: WPY 500/KB
Max capacity 510kg
Readability
2g (for mass standards 200kg)
5g (for mass standards 500kg)
Repeatability *
2g (for mass standards 200kg)
4g (for mass standards 500kg)
Electric compensation range 0 ÷ 510 kg
Supplementary weights none
Stabilization time 120 s
Adjustment / calibration External
Power supply 110 ÷ 230V AC / 50 ÷ 60Hz / 13,5 ÷ 16V DC / 1,1A
Pan size
φ1000 mm
Weighing unit dimensions 1347x1166x200 mm
Control unit dimensions 206x140x70 mm
Net weight 200 kg
Ambient conditions:
Working temperature +15°C ÷ +35°C
Change rate of working temperatures
±0,5°C/h
Atmospheric humidity 30% ÷ 70%
Change rate of atmospheric humidity
±10%/4h
Technical data: WPY 1000/KB
Max capacity 1100kg
Readability
5g (for mass standards 500kg)
10g (for mass standards 1000kg)
Repeatability *
4g (for mass standards 500kg)
8g (for mass standards 1000kg)
Electric compensation range 0 ÷ 1100 kg
Supplementary weights none
Stabilization time 120 s
Adjustment / calibration External
Power supply 110 ÷ 230V AC / 50 ÷ 60Hz / 13,5 ÷ 16V DC / 1,1A
Pan size
φ1000 mm
Weighing unit dimensions 1347x1166x200 mm
Control unit dimensions 206x140x70 mm
Net weight 200 kg
Ambient conditions:
Working temperature +15°C ÷ +35°C
Change rate of working temperatures
±0,5°C/h
Atmospheric humidity 30% ÷ 70%
Change rate of atmospheric humidity
±10%/4h
* Repeatability is expressed as a standard deviation of six ABBA cycles
(according to R111 OIM) in stable laboratory conditions.