User manual
100
Example: 
S I CR LF – computer command 
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command done, mass value 
in basic unit is returned immediately. 
28.4.7. Send immediate results from all platforms in basic units 
Syntax: SIA CR LF 
Possible answers: 
SIA_I CR LF 
- command comprehended but cannot be executed at the moment 
MASS FRAME „P1” CR LF 
MASS FRAME „P2” CR LF 
- mass values are immediately returned from all 
 platforms in basic units 
Frame format with mass from subsequent platforms as indicator reply: 
1 2  3  4  5  6  7-15  16  17 18 19 20 21 
P n space  stability  space  sign mass space  unit  CR LF 
n  - weighing platform number 
mass - 9 characters justified to the right 
unit  - 3 characters justified to the left 
Example:
Let us assume that both platforms are connected to indicator PUE HY. 
S I A CR LF – computer command 
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF 
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF  - command done, mass values 
from both platforms are returned in basic units 
28.4.8. Send the stable result in current unit 
Syntax: SU CR LF 