User manual

- 75 -
WPT
60H5
WPT
150H5
WPT
300H5
WPT
150H6
WPT
300H6
Scale type:
WPT
60H5/K
WPT
150H5/K
WPT
300H5/K
WPT
150H6/K
WPT
300H6/K
Max capacity 60kg 150kg 300kg 150kg 300kg
Min capacity 20g 50g 100g 50g 100g
Tare range -60kg -150kg -300kg -150kg -300kg
Platform size 600 × 600mm 800 × 800mm
Work temperature -10°C to +40°C
Ingress protection rating IP 67 - platform, IP 66/67 - indicator
Power supply
230V AC, 50Hz / 11V AC
and internal gell cell SLA 6V accumulator
Supplied from batteries Average operation when supplied from batteries 45h
Display LCD with backlight
22. TROUBLE SHOOTING
Problem Cause Solution
Discharged batteries.
Connect to mains or
change batteries
Turning on does not
work
No batteries (not installed or
improperly installed)
Check the correctness of
installation (polarization)
The scale turns off
automatically
„t1” set to „YES” (Power
save)
In „othr” submenu change
„5.4 t1” to „no”
After turning on „LH”
message on the display
Loaded weight pan during
powering up
Unload the pan. Then the
scale will indicator zero.