User Manual

- 44 -
Return to weighing:
Press the key twice.
16.5.2. Mass of standard inscribed to scale memory
Procedure:
Enter to <PrcB> function:
The program goes to the weight display window:
Using and set standard mass,
where: - digit selection, - digit setting.
Confirm the entered value by pressing ,
You will see the indication equal to 0,000%,
From this moment display will not indicate the mass of weighed
load but deviation of the load mass placed on the pan referring
mass of standard (in %).