User Manual

- 28 -
Pr_n
noStAb
-
immediate printout
(not accessible in verified scales)
Pr_n
StAb
-
sending stable results
Pr_n
rEPL
-
automatic operation
Pr_n
CntA
-
continuous transmission in basic unit
Pr_n
Cntb
-
continuous transmission in present unit
Return to weighing:
see 11.2.2.
14.2. Minimal mass threshold
This function is necessary while working with automatic tare
or automatic operation or weighing animals.
Automatic tarring will not be applied until the indication (gross)
is lower than the value inscribed in S_Lo parameter.
In automatic operation measurements (net) are sent via RS232
when the indication is equal or greater than the value inscribed
in S_Lo parameter.
Weighings animals is performer when the indication is equal
or greater than the value inscribed in S_Lo parameter.
Procedure:
Enter the submenu <P2.Prnt> and then: