34.1.3. Weighing platform connecting
• 6-wire strain gauge sensor platform connection
6-wire strain gauge sensor can be connected to weighing module board as
instructed below:
WEIGHING MODULE
SIGNAL
STRAIN GAUGE
SENSOR SIGNAL
NOTICE
E SCREEN
(according to the rule
of connecting shields)
REF+ SENSE + JP1 not soldered
REF- SENSE - JP2 not soldered
IN+ OUTPUT+
IN- OUTPUT-
+5V INPUT+
AGND INPUT-
• 4-wire load cell cables
For 4-wire load cell cables look at the drawing below:
- 119 -