34.1.1. Module technical specification
Maximal number of convertor units 8 388 608
OIML class III
Number of verification units 6000e
Maximal signal gain 19mV
Maximal voltage per verification unit 3,3 µV
Minimal voltage per verification unit 1µV
Minimal strain gauge impedance 90
Maximal strain gauge impedance 1200
Strain gauge converter supply voltage 5V
Strain gauge converters connection 4 or 6 wires + shield
34.1.2. Weighing platform signal wires colours
Strain gauge
sensor
determiners
Colours
according to the
RADWAG
standard
W
eighing module
determiners
+INPUT brown +5V
-INPUT green AGND
+OUTPUT yellow +IN
- OUTPUT white -IN
+SENSE gray +REF
- SENSE pink - REF
SCREEN yellow-green
(according to the rule
of connecting shields)