User Manual Scales of TMX series Manual number: ITKU-36-06-09-11-A Terminal E2R Ewidencja v. 1.1.6.119 MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG 26 – 600 Radom 28 Bracka Street - POLAND Phone +48 48 38 48 800, phone/fax. +48 48 385 00 10 Selling department +48 48 366 80 06 www.radwag.
SEPTEMBER 2011 -2-
TABLE OF CONTENTS 1. 2. 3. 4. 5. 6. 7. 8. 9. INTENDED USE ....................................................................................................................6 PRECAUTIONS .....................................................................................................................6 WARRANTY CONDITIONS...................................................................................................6 MAIN DIMENSIONS ....................................................................
21. TRANSACTIONS ...............................................................................................................46 21.1. Starting working mode................................................................................................46 21.2. Starting a transaction..................................................................................................47 21.3. Continuing a transaction.............................................................................................
29.8. Closing WeighingServer program.............................................................................107 30. ERROR MESSAGES .......................................................................................................107 31. LABEL DESIGNING ........................................................................................................107 31.1. Label pattern making ................................................................................................108 31.2.
1. INTENDED USE Scales of TMS series with TERMINAL E2R EWIDENCJA are intended to be used for measuring and at the same time recording the data in local database or on server. The application is especially important when the weighed products are not directly related one to another by any common data or they are not a part of the weighing process with its beginning and end, for example deal or order.
D. Warranty does not cover: • Mechanical failures caused by inappropriate maintenance of the device or failures of thermal of chemical origin or caused by atmospheric discharge, overvoltage in mains or other random events, • Cleaning service. E. Warranty expires after: • Access by an unauthorized service, • Intrusion into mechanical or electronic construction of unauthorized people, • Removing or destroying protection stickers. F. The detailed warranty conditions are specified in warranty certificate. G.
5. DESCRIPTON OF CONNECTORS Back side view of PUE 5 Connectors in standard version: 1 2 3 4 5 6 7 8 Strain gauge cable gland USB connector USB connector RS232, RS485 connector Ethernet connector Power cord cable ON/OFF switch Glands of optional equipment (8IN/8OUT, 4IN/4OUT, RS485 via gland, additional platform, etc) NOTICE: All connectors can be used in any configurations.
6. UNPACKING AND ASSEMBLY A. Remove the scale from the packaging, B. Place the scale on even and hard surface far from heat sources, C. Scale should be levelled by turning levelling foot. Levelling is correct if air bubble is in central position of level indicator located on scale’s base. 7. GETTING STARTED • Turn on the scale pressing the ON/OFF switch on the back side of the terminal. Then Windows loading will start.
8. LOGGING ON You have to log on every time you start the scale.
Caution: „Operator 1” has got the „administrator’s” authority, what makes a scale’s user have direct access to all the program functions. The application consists of four authority levels, which are described later in the manual. In the information line the following can be found: Log off time - Time of user’s logging off or turning on terminal’s applications User’s name Time of logging - User’s name and the time of logging 9. SCALE WINDOW DISPLAY 9.1.
9.2. Buttons’ functions Zeroing of scale Tarring Hand-operated tare entering 10. WEIGHING PROCEDURE appears the result can be read. Put a load on the scale pan. When In order to record the weighing result press: Caution: To save the weighing result operator must be logged on and product must be chosen from database first.
• Do not load the pan with concentrated force: • Avoid side loads, particularly side shocks should be avoided 10.1. Tarring In order to get the net weight put the load on pan and press when it is stable ( zero is indicated and Net symbol appears in top left corner of weighing window). After placing a load on the weight pan net mass will be shown. Tarring is possible within the whole range of the scale. While tarring, remember that the maximum capacity of scale should not be exceeded.
After unloading product and packing from the pan the display shows the tarred value with “minus” sign: Notice: Tarring process cannot be performed while the display shows a negative or zero value. Then Err3 will appear on weighing window display. 10.2. Hand operated tare value entering It is also possible to enter a tare value by hand.
Notice: Zeroing is possible only within the range of ±2% of maximum scale’s capacity. If the zero value is beyond this range, Err2 message will be displayed. 10.4. Weighing on two range scales Switching between the I range and the II range weighing appears automatically without operator’s presence when I range maximum value is symbol in the top exceeded. Starting II range weighing is indicated by left corner of the scale window. When loading is eliminated the indication returns to zero value.
1 2 3 4 5 6 7 8 - Information line Recorded weighing results window Mass bar chart Database of statistics Log off/log on button Weight window Choosing items from database buttons Programmable buttons Choice window description/edition of database items 9 8 1 7 2 6 3 4 1 2 3 4 5 6 7 8 9 - 5 Displayed items filtering fields Database items displaying window Sorting buttons Ten latest displayed items button Addition/edition/elimination database items button Change approval/refusal or selection button
Screen keyboard window - Window clearing - Cursor moving arrows 12. PRODUCT CHOICE In the main program window the scale’s user can choose product from database by name or code, as the picture shows: Choosing by code Choosing by name To remove the selected product press . 12.1.
• Choose the demanded product, then the program will automatically return to main window, where the product’s code and name will appear in appropriate fields: • In case of large number of products in the database, user can search: - by code - by name - by EAN code of product As it is shown in the picture below: - searching by code - searching by name - searching by EAN code In order to eliminate filtering of code, name and EAN fields, press button.
12.2. Choosing a product by code Procedure: • Enter the product choosing by code mode according to point 12 of the manual so you can see : • Using the screen keyboard load the demanded product’s code and accept it pressing button. • The program will automatically return to the main window, displaying code and name of the chosen product in appropriate fields. 13.
CAUTION! Contractor’s choice window will be accessible after setting up its profile in the main application window, according to point 27.5.1 of the manual. 13.1.
As it is shown in the picture below: - Searching for contractor by code - Searching for contractor by name In order to eliminate code and name fields’ filtering, press button. 13.2. Contractor choosing by code Procedure: • Enter constructor choosing by code mode according to point 13 of the manual, so you can see the following: • Using screen keyboard, load the required contractor’s code, then accept it pressing button.
14. CHOOSING A PACKING In the main application window the user can choose a packing by its code or name as follows: Choosing by code Choosing by name To remove a package press . Caution! Packing choosing window will be accessible after setting up its profile in the main application window, according to point 27.5.1 of the manual. 14.1.
• Press required packing, then the program will return to the main window displaying code and name of the chosen packing • In case of large number of packages in database user can filter or choose by: - code of packing, - name of packing, As the picture shows: - Searching packages database by code - Searching packages database by name In order to eliminate code and name fields’ filter, press button. 14.2.
• Using screen keyboard, enter the demanded code of packing and accept it pressing • button The program will automatically return to the main window displaying code and name of chosen packing in appropriate fields. 15. STORE CHOOSING User can choose a source and target store according to following picture - Target store choosing - Source store choosing To remove a store/warehouse press .
• Choose demanded store, then the program will automatically return to main window, displaying name of chosen store in appropriate field: • In case of large number of stores in database, user can filter or search by: − store’s code − store’s name, As the picture shows: - Searching stores’ database by code - Searching stores’ database by name In order to eliminate code and name fields’ filter, press - 25 - button.
16. LOT CHOOSING Scale’s user can load a lot number for each weighing. In order to do that, Press the button appropriate for loading a lot number in the main application window. Lot/batch window can be cleared by pressing . Caution! Lot choosing window will be accessible after setting up its profile in main application window, according to point 27.5.1 of the manual.
17. LOT 2 CHOOSING User weight for each weighing can indicate the number of LOT 2. Procedure for defining the LOT is the same as in paragraph 16 of the instructions. 18. QUANTITY NUMBER CHOOSING Procedure: In order to assign a weighting number which acts as a descriptive panel, press the main weight of responsibility for the introduction of a quantity. Then, with the numeric keypad enter any number. Entered number is displayed on the button and is stored in the weighing .
19. DATABASES 19.1. Access to databases edition Procedure: • While in database choice window, press following will be displayed: button, then the 19.2. Operators’ database Caution! Operators’ database edition is accessible in the program’s options for logged user who has the administrator’s or advanced authority. Operators’ database can be also edited while user’s change-logging. Procedure: • In order to edit operators’ database, press • Press • in operator’s log off window.
• Access to operators’ database window will be displayed: 19.2.1. Adding an operator Procedure: • Enter the operators choosing window according to point 19.2 of the manual, then press: according to point 19.
And as the picture shows: name* code* authority set up password - operator’s name field - operator’s code field - user’s authority type - setting up user’s password *) - filling in is obligatory • Press text field or button (code, name) in order to load operator’s data The buttons’ functions: - Approval of loaded new information - Refusal of loaded new information - 30 -
• After operator’s code and name loading, set up authority level pressing a button below; choose an option from unrolled list: . • Set up operator’s password, pressing Using keyboard that appears on the terminal’s screen, enter the password and accept it. Caution! You have to enter identical password twice, otherwise information about error will be displayed: • In order to accept adding an operator press: , • In order to refuse adding an operator press: . 19.2.2.
this is what will appear on the screen: Complete the fields according to information in point 19.2.1 of the manual • In order to accept all changes, press • In order to refuse changes, press , . 19.2.3. Eliminating an operator Procedure: • Enter operators choosing window according to point 19.2 of the manual, then according to point 19.1 press: then the window will appear: Accept eliminating an operator pressing: - 32 - .
19.2.4.
19.3. Base of products 19.3.1. Adding products Procedure: • Using product choosing by name button in main program window enter base of products, then, according to point 19.1 of the manual, press: and this will appear: • press: button or text field in order to load data. • in order to accept adding a product press: • in order to refuse changes press: - 34 - , .
Fields to be completed: Name* Code* EAN Price Mass Tare Min Max Validity time Temperature Edit Label Field VAT [%] Label - name of product code of product EAN bar code of product retail price of product net weight tare weight product’s min threshold product’s max threshold No of validity days of a product after the day of - weighing - Storage temperature (text field) - This field can hold any information about the product - Interest rate of VAT - File selection box printed labels for the goods *) – fil
After pressing a window appears in which additional information can be inscribed e.g.: ingredients. Every line (L1-Ln) can be printed separately, see additional information in ch. 31.2. and then enter the text for the In order to add this field press line. The line is displayed in the window.
Functions of buttons in the window: - Add a line - Edit a line - Delete a line - Change line position - Cancel changes Confirm changes 19.3.2. Product’s editing Procedure: • Using choosing product by name button in the main program window enter base of products, then choose product to be edited, and according to point 19.
Fill in the fields according to point 19.3.1 of the manual • In order to accept changes, press • In order to refuse changes, press , . 19.3.3. Product eliminating Procedure: • Using product choosing by name button in the main program window enter base of products, choose product to be eliminated, and then according to point 19.1 of the manual press: chosen product will be eliminated. 19.4. Contractors’ base 19.4.1.
the following will appear: • Press • button or text field in order to load data. , In order to accept adding a contractor press: • In order to eliminate changes press .
19.4.2. Contractor’s edition Procedure: • Using contractor choosing by name button in main program window, enter contractors base, choose a contractor to be edited, then according to point 19.1 of the manual press: then the following will be displayed: Complete the fields according to point 19.4.1 of the manual. • In order to accept changes press , • In order to refuse changes press .
19.4.3. Contractor’s elimination Procedure: • Using contractor choosing by name button in the main program window, enter contractors’ base, choose a contractor to be eliminated and according to point 19.1 of the manual press: chosen contractor will be eliminated from database. 19.5. Base of packages 19.5.1. Adding a packing Procedure: • Using packing choosing by name button in the main program window, enter base of packing, then according to point 19.
• In order to accept adding a packing press: • In order to refuse changes press: , . The spaces’ meaning: Name* Code* Barcode Mass [kg] - name of packing code of packing bar code of packing weight of packing *) - fill in obligatory 19.5.2. Packing’s edition Procedure: • Using choosing a contractor by name button in the main program window, enter contractors database, choose a contractor to be edited, then , according to point 19.
complete the fields according to point 19.5.1 of the manual. • In order to accept changes, press: , • In order to refuse changes, press: . 19.5.3. Eliminating a packing Procedure: • Using choosing a packing by name button in the main program window, enter packages base, choose a packing to be eliminated and, according to point 19.1 of the manual, press: chosen packing will be eliminated from database. 20.
In the interface window tick Counting Pieces. You will see . In the interface window, select the Operating modes and Counting pieces tab, select Active. • Press 2 times. • The introduced changes are confirmed by pressing The working mode is on when the pcs unit is on the display: In the weighing window additional column Pieces appears apart from weighing in g.
20.2. Setting reference unit mass The reference unit mass can be set by estimating single piece mass in the main program window using a function key attributed to this function. Settings are described in ch. 27.5.2 of this manual. After zeroing the platform put pieces on the pan and inscribe the quantity in field Number of Pieces. Single piece mass is estimated automatically. The reference unit mass can be also set in the assortment database.
The third way of setting the reference mass is to inscribe its value in the field Mass in the window above. 21. TRANSACTIONS PUE5 weighing terminals with the extended software version can operate in working mode transactions. The weighing process is connected with selling transactions – purchasing, production orders, servicing orders and stock management. A new transactions can be created on the terminal. It can be temporarily suspended or terminated.
• Press two times. • Confirm the introduced changes by pressing 21.2. Starting a transaction There is button new in the main window to start a transaction. Choose transaction type.
Types of transaction: M-M Transaction Between Warehouses - Moving a product from one warehouse to another. SPR Sell - Transaction of selling that requires outlining of a contractor and a source warehouse. ZAK Purchase - Transaction of buying that requires outlining of a contractor and a target warehouse. The next step is to choose a contractor to be attributed to the transaction and a source or target magazine. When moving between warehouses contractors are not necessary.
Accessible warehouses: Code 1 - Sell Code 2 - Purchase Code 3 - Warehouse1 After choosing all necessary parameters in the main program window detailed transaction information appears and a transaction symbol is attributed. All weighings in the transaction are attributed to a given unique symbol. A transaction in progress can be terminated or completed using an appropriate button break or end. Suspending a transaction allows to postpone it until it is selected again.
From the list of transactions choose the one you want to continue and press Ok. Until the transaction is closed it is accessible on the list of suspended transactions. 22. STATISTICS All the statistic data is regularly updated after loading each measuring data to scale’s database. Data interlock can be seen in bottom left corner of the main program window. Statistic data is updated globally irrespective of weighed product.
23. WEIGHING RECORDING All weighing results are saved in MS SQL database, which is stored locally or on another weighing terminal or server. After pressing SAVE button, weighing result will be displayed automatically in the chart. The last twenty weighings are present on the list. All weighings are displayed in grams in the table. Additionally, user can sort certain columns increasingly or decreasingly, pressing chosen column’s name.
The primary filter, the filter weights is presented as of the date - to date. By default, when you start windows are presented last two days. After selecting the weighing and selecting edit and delete, the weighting will be awarded as an archive which means that it will not be presented in the report. Jeżeli ważenie zostało odznaczone przypadkiem jako archiwalne to za pomocą funkcji edytuj i dodaj, możemy wyłączyć status archiwalne i od tej pory ważenie będzie prezentowane w raportach. 24.
25. LOGGING OFF Log off button in the main program window is used for logging off and switching off the terminal. The options are accessible if the logged user has got administrator’s or advanced authority. To exit the operator’s log off window, Press: 25.1. Logging off Log off function is used when a scale’s user is finishing his work and the terminal is not switched off.
• Then press Change-log button: • List of accessible users will be displayed; after one of users is chosen, the program returns to its main window automatically. 25.3. Switching off a terminal User can switch off a weighing terminal only if he is logged on.
26. CHECKWEIGHING THRESHOLDS In the main program’s window user can see checkweighing thresholds’ results (MIN,MAX) in a bar chart Checkweighing thresholds data is taken from database of a product or programmed with function buttons described in point 27.5.2 of the manual. 27. PROGRAM’S OPTIONS Procedure: • In order to enter program’s options, user with administrator’s or advanced authority must be logged on.
• Window with weighing parameters will be displayed automatically Weighing Server program configuration platforms connected by weighing module MW02 is described in point 24 of the manual. 27.2. Operators The option enables operators’ edition and setting logging and authorization mode.
Then the window with accessible operator’s options will be displayed: 27.2.1. Operator’s edition Edition, that is adding, changing or eliminating an operator has been described in point 19.2 of the manual. 27.2.2. Log on procedure Procedure: • Enter operator’s options as instructed in point 27.
• Choose logging procedure: Accessible option will be marked with . • in order to accept loaded changes, press: • in order to refuse changes, press: , .
27.2.3. Authorization Procedure: • Enter operator’s options as instructed in point 27.2 of the manual, then press: • Choose authorization mode: • In order to accept changes, press: , • .
Note that: Without authorization - In order to log on to a program, the user is supposed to enter chosen operator’s password Password - While logging, password entering option is not active CAUTION! If Without authorization and Select from list point 27.2.2 options are accessible at the same time, then user with subordinated authority can log on one of administrator’s accessible profiles, which enables him to get access to most of program’s options. 27.3.
• Accessible devices will be displayed in the window below: 27.3.1. Scales The option enables configuration of scales connected to the weighing terminal PUE5 with Ethernet or RS232/485. Procedure: • Enter devices option as instructed in point 27.
• Choose a scale to be edited then press: .
Note that: Description - optional characteristics of scale Host - Port - scale’s port’s number UDP/TCP Database id - Platform’s identification number LO Threshold - LO threshold value for a particular platform. host’s IP address, default 127.0.0.1 specifies local computer’s address, LO Threshold parameter is related to the function of the automatic operation and control outputs.
• Then window of options accessible for designing and labelling printer prints setting will be displayed • In order to accept changes press: , • In order to refuse changes, press: . Note that: Name Design labels Label patterns file Active label printing Here you can choose labelling printer accessible in operational system. After printer has been installed in operational system, it will be added to the list automatically. Starting Edytor Etykiet program, used for label - designing.
Labels printing active while measurement recording Additional options associated with printing labels, which allows you to disable printing of labels during the recording. Enable option to print labels for each record. Printed labels Number of printed labels, - When set to 0, you write the part number printed labels. CAUTION! Printing labels is available after setting the printer, the standard label and select the two options Active printing.
27.3.3. CGM – Apparatus for testing conformation The program works with the apparatus to examine the conformation which allows registration at the weighing of information on meat quality. Hardware configuration requires you to set the communication port and speed. Procedure: • Enter devices option as instructed in point 27.3 of the manual, then press: • You will then see a dialog where you can set the COM port to which the terminal device is connected.
27.3.4. Output mode Output mode options allows you to configure the two selected outputs are activated depending on the current position in relation to 0, the threshold MIN and MAX currently selected item. This feature can be used to control the process of dispensing the goods or alarm thresholds. Procedure: • Enter devices option as instructed in point 27.3 of the manual, then press: • Then in the window, select one of the modes for thresholds and assign the starting threshold.
OUTPUT MODE THRESHOLD Threshold 1,2 to MIN Threshold 2 to MAX Threshold 1 to MIN Threshold 1,2 to MAX Threshold 1 to MIN Threshold 2 to MAX 0 Threshold 1 from MIN Threshold 1,2 from MAX 0 Threshold 1 from MIN Threshold 2 from MAX 1 2 1 2 1 2 1 2 3 1 2 3 STATE OF MASS 0 - MIN MIN- MAX 0 - MIN MIN- MAX 0 - MIN MIN- MAX 0 - MIN MIN- MAX > MAX 0 - MIN MIN- MAX > MAX Dosing rough 1 0 1 1 1 0 0 1 1 0 1 0 Dosing accurate 1 1 0 1 0 1 0 0 1 0 0 1 CAUTION ! In the case of mode 3 release trigger outputs to turn
27.4. Reports The option enables preparing and printing the weighing results report or write a report on the terminal disk in a file format PDF, XLS or CSV. NOTE: Versions of reports are described below 9.11.9.0. PUE5Reports.dll file reports is located at Terminal installed E2R. Procedure: • Enter the programme’s options as instructed in point 27 of the manual, then press: • Then the following window will be displayed.
27.4.1. Date Date is the main criterion of report preparing. Procedure: • Enter reports window as instructed in point 27.4 of the manual • Set up the initial and final calendar date, using date field. button placed near Calendar window will be closed after day has been chosen.
27.4.2. Laps Using laps( Report 1, Report 2 ) located on the left, user sets various report displaying options. Following options can be used with laps column displaying order change column adding and eliminating from the view text field column filtering button column filtering increasing and decreasing column sorting After exiting the programme, laps remain unchanged until return to default view function is used.
• Menu In order to get access to extended menu, you need to hold your finger a bit longer on a column.
Full Collapse * Clear Grouping * - winding up all the results elimination of grouping accessible after choosing a grouping field *) – options accessible if grouping field window is switched on • Columns filtering In order to filter a column’s data, write down the needed phrase in first line: . Mark the first line and press: After having entered needed phrase, press: , Records similar to the given phrase will be displayed in the column.
In order to clear up loaded data displaying column filter, press in main window. • Sorting Sorting is possible after column or menu pressing . Sorting mode is . displayed as changing sign: • Choosing columns to customization window of the drag on two Use the button sides of any column.
Mass Gram Masa unit Unit Operator Product Contractor Tara LOT Platform Destination Max Min Pieces Product code Transaction Source - Mass of weighing in gram unit Mass unit set in the base of the product Unit set up in the database of the product Operator Name Product name Name of contractor Tara LOT, batch symbol Number of platform The name of the target store Threshold for the maximum weighted product Threshold for the minimum weighted product Number of pieces Product code Symbol transactions Name of the
After having prepared appropriate report, press: .
zoom in print monitoring closing zoom in in order to see enlarged part of the report closing the report print monitoring PDF File - Export to PDF file format Excel File - Export to XLS file format CSV File - Export to CSV file format Variable name file export - Variable export file name - Constant export file name 27.4.4. Export to a file Adequately prepared the report print to a printer installed in the operating system.
Procedure: • Enter program’s options as instructed in point 27 of the manual, then press: • Choose demanded option from Others window 27.5.1. Interface view User can modify main program’s window, that is, turning on/turning off certain elements which can be seen in the main window.
Fields of the main screen: − contractor choosing button − packing choosing button − target store button − source store button − lot number loading button, − lot 2 number loading button, − entered quantity loading button, Elements of the main screen: − statistics windows, − bar chart windows, Operating mode of the main screen: − transactions, − counting pieces, − CGM, − Orders.
• Choose elements (functions) of main program window from interface view window. • In order to accept changes press: , • In order to refuse changes, press: . 27.5.1.1. Choice fileds Buttons in the column Required select the required parameters for the right of the main window while writing manual or automatic. This is equivalent to the need to fill some fields in the main window. Buttons in the column Visible meet accessibility (visibility) of the button in the main window.
27.5.1.2. Operating mode Tab transactions module provides transaction activation button in the main terminal and the function of enforcing the window displays the number of record labels each time you weigh.
Counting pieces Tab provides activation counting module in the main terminal. Tab CGM provides activation of cooperation with the camera module to study conformation in the main terminal. CGM option to remove the data after saving measure allows for the cleaning of the information in the main window after recording. Optional data entry required for measuring input forces from the camera when saving measure.
Orders tab allows you to order, including the module which is an extension module for transaction orders sent from the E2R transactions from your PC. 27.5.1.3. Elements On the elements of statistics and activate the panel bar graph, in the main window.
27.5.1.4. Response On the behavior can activate the functions performed by the program. Tarring after saving allows you to automatically tare the platform after each measurement record. Ask to offset the suitability of the goods during the recording allows you to display a window during the recording in which the operator can extend the product's shelf. To the days of the goods listed in the database will be added to the number of days entered in the window.
27.5.2. Buttons’ functions User is able to configure main program’s window programmable (functional) buttons. Procedure: • Enter Others option as instructed in point 27.5 of the manual, then press: • From Buttons’ Functions window choose visibility of main program’s window buttons.
• After choosing button’s visibility, choose adequate function: • In function choosing window, choose one of accessible functions: - 86 -
• To accept the changes, press: , • To refuse the changes, press: . • After having chosen button’s function, enter button’s label. • In order to accept changes press: , • In order to refuse changes, press: .
Accessible functions list: − − − − − − − − − − − − − − − − − − No statistics clearing, Record of printing a single label, Record label without printing, Operator’s logging off, Operator’s logging on, Mass pattern definig Closing program, Increasing the precision, Reports, Tare loading, Lot number entering, Measurement saving, minimum threshold setting, maximum threshold setting creating a transaction, terminating a transaction. Label print 27.5.3.
• Choose required language and accept it pressing: , CAUTION! Changing program’s language version does not influence Weighing Server (weighing parameters window) or Label Editor (label designing) language versions. 27.5.4. Application closing Program’s closing and passing to Windows XP Embedded operational system are accessible in other options. Procedure: • Enter Others option as instructed in point 27.
• Accept the choice in application closing window. 27.6. Alibi E2R Terminal has the ability to save records of weighings in a place independent of the SQL database. Configuration options are available for the administrator and power user has access to view content. Weighing saved in the Alibi include: • Time, • Mass, • Tare, • Unit, • Product name, • Operator name, • Contractor name, • Serial number, • Source stock, • Target stock, • Packaging. Procedure: • Enter Others option as instructed in point 27.
• In the full terminal window will appear weighing Alibi memory. • Using the Actions you access to options related to memory Alibi.
27.6.1. Enabling write memory Alibi Enabling the Alibi is pressing the select button, Operations and Active record measurements as described below. The importance of the buttons: - Function not acrive - Function active 27.6.2. Deleting a measurement of memory Alibi Button to clear the complete measurements, measurements removes the set has secured days. The program allows you to record multiple measurements with the number is only limited at the time of removal.
27.6.3. Export measurements alibi to a csv file All listed measurements are saved to a file in *. csv separated by semicolons. Each time a record will be confirmed following message. 27.6.4. Filter In the main window of measurements stored in the Alibi is available filtering and grouping of records. To this end, hit the button , and then in the window below the box indicate that you want to search and view the memory window Alibi.
27.7. IN/OUT configuration For terminal equipped PUE5 derived I / O configure their operation. Procedure: • Enter Others option as instructed in point 27.5 of the manual, then press: 27.7.1. Inputs configuration On the input we can assign a specific function to the selected input depending on its activation or deactivation. The assignment of the input is activated by pressing work or until you reach the input signal.
The list of available functions assigned to the input: • Disabled, • Tarring, • Zeroing, • Increased accuracy, • Print. 27.7.2. Outputs configuration On the output assign specific functions activating or inactivating the selected output. The assignment of the output activation will produce a signal at the output. The assignment of the deactivation of the output will disable the output.
The list of available functions assigned to the input: • Disabled • Mass first stable > LO • Mass below LO • Mast last Stable > LO • Mass over LO • After saving measure • Stable • No Stable 27.7.3. Other options On the other options are available concerning the terms of inputs / outputs that can make the operator from having to log in, or completion of all required data.
28. CONFIGURATOR PROGRAM 28.1. Basic Configuration Terminal to Terminal PUE5 E2R records that has the settings allow you to work in the local SQL database. If the terminal is designed for networking, need to change the factory settings. To run the setup file Config.e2r delete the folder C:\Program Files\RADWAG\Terminal E2R Ewidencja in the file that contains all the necessary terminal settings. In the following startup screen, press - 97 - .
The next window allows you to enter server parameters and settings. Where: Server - Database Login Password - Name or IP of the server and SQL instance E2R SYSTEM works for instance BAZARADWAG The name of the database, by default E2R Login to the SQL database, by default sa The password for the SQL database, by default radwag Button for checking the connection to the set parameters of the database. Window proper connection to the database.
The window definition ID for terminal, which is unique in the case of terminal operation in the system E2R. Once fully configured, restart the program. 28.2. External configuration In parallel with basic configurator that runs when the file is missing Config.e2r in the program folder. EwPue5Settings.exe program is available in the program folder.
In the configuration, get access to change the parameters stored in a config.e2r - number, weight and data on the SQL database. Window settings SQL database. Where: Serwer - Baza danych UserID Hasło - Name or IP of the server and SQL instance E2R SYSTEM works for instance BAZARADWAG The name of the database, by default E2R Login to the SQL database, by default sa The password for the SQL database, by default radwag The settings window of the terminal.
Options for loading and saving configuration can be found in the Configuration menu. Where: Wczytaj aktywną Wczytaj bieżącą - Wczytaj z pliku Zapisz aktywną - Zapisz bieżącą - Zapisz jako - Load configuration from file Config.e2r Load re-opened for the configuration of the (active) file Load configuration from the selected file Writing configuration settings to a file Config.
29. WEIGHING PARAMETRES SETTING User can change basic weighing setting using Weighing Server program. 29.1. Weighing Server program starting Start scale software as instructed in point 27.
29.2. List of software menu Hold finger for about 2 s on scale software window, menu list of accessible software will be displayed: 29.3. Parameters of scale software 29.3.1. Readout of parameters To read out scale parameters press: indicate parameters list accessible for edition: , scale software will 29.3.2. Save changes procedure After having changed the parameters, save the values by pressing: .
29.4. Setting a filtering level Procedure: • Readout parameters according to point 29.3.1 of the manual, • Press and rewind Filter parameter value window. After selecting required parameter value save changes according to point 29.3.2 of the manual. Notice: The higher filtering level the longer scale stabilization time. 29.5. Median filter This filter eliminates short pulse disturbances (for example: mechanical shocks). Procedure: • Readout parameters value according to point 29.3.
After selecting the required value save changes according to 29.3.2 of the manual. 29.6. Autozero function The autozero function has been introduced in order to assure precise scale’s indications. This function controls and corrects „0” indication. While the function is active it compares the results successively with constant frequency. If results differ less than the declared value of autozero range, for example 1unit, the scale will be automatically zeroed and the and zero indicator will be displayed.
After selecting the required parameter value save changes according to 29.3.2 of the manual. 29.7. Scale software settings After starting program menu list (see point 29.2 of manual) press with finger option : Scale software will display settings window: Accept change of parameters value with key: - 106 - .
Notice: Additional weighing platform module MW02 is working on port COM2. Languages available: • Auto • • • Polish English German - Setting language version compatible with the version of the operating system. Set the address track to the software keyboard on: C:\Program Files\screen keyboard\KEYBOARD_mini.exe 29.8. Closing WeighingServer program In order to exit the program press the button: window. on the main 30.
Example: Setting in a scale a pattern adequate for the following label: 31.1. Label pattern making Procedure: 1. Start EDYTOR ETYKIET R01 computer program as instructed in point 29.3.2. or operational system according to the following description: • • • • • press START on the bottom task bar indicate Programs option, choose RADWAG option, choose Edytor Etykiet R01, click Edytor Etykiet R01 symbol. 2. While starting, the following window will appear: 3.
4. In order to add text to a label, click following window will be displayed 5. Start the keyboard pressing: screen. on the left task bar, then located in top right part of the 6. In space enter the text: Meat Plant „MEATON”, accept with and place it on label using appropriate type size.
7. Click again on and in the window „Enter text” insert {4}, and confirm by pressing and place the required variable in the right place on the label, 8. Place other constant texts and demanded variables on the label as instructed above. 9. Click and in „Load graphics” window rewind list: 10. Choose required <*.bmp>, accept with appropriate label place and put graphics in Caution: <*.bmp>located on label will be printed after the graphics have been sent to printer’s memory.
12. Using keyboard enter variable in space {52}. 13. Accept changes with place. and put a code in appropriate label 14. Save ready label pattern choosing , then lab. 15. Set up a label in Terminal E2R Evidence application as instructed in point 27.3.2. of the manual. Caution: Saved label pattern with*.lb extension is not editable file. Therefore additional label pattern recording with *.
{31} {35} {48} {49} {50} {51} {52} {53} {54} {55} {56} {57} {59} {60} {62} {64} {65} {66} Platform number Counting pieced: unit mass in calibration unit Product: Temperature Product: Description Product: Name Product: Code Product: EAN code Product: Mass Product: Tare Product: Unit price Product: Minimum Product: Maximum Product: No of validity days Product: VAT Product: Expiry date Product: ingredients or any other additional information Contractor: Name Contractor: Code {67} Contractor: Tax ID {68} C
In counting pieces variable {6} „Net mass in present unit” presents the number of pieces {10} the current unit is “pcs”. In case of variable {64}, each line (L1-Ln) is formatted according to the pattern: Line 1 – {64:L1}, Line 2 {64:L2}, etc. 32. DIAGRAMS OF CONNECTION CABLES The scale in STANDARD version can cooperate with: • • • computers slip printers KAFKA, EPSON, KYOLINE label printers CITIZEN, ZEBRA, 32.1. USB cable (adapter) USB cable (adapter) PT0084 32.2.
32.3. RS232 printer cable RS232 printer (CITIZEN, EPSON) cable PT0019. 32.4. Ethernet cable Ethernet cable P0198: 32.5.
32.6. RS232, RS485 cable – colours Colours for “M12” standard cables. The figure gives an example of the type of cable. 32.7. RS232C cable – colours Colours for “M12” standard cables. The figure gives an example of the type of cable. 33. CONNECTORS Caution: Depending on the number of installed additional modules the number and location of glands and connectors can change. Standard option connectors and glands’ location does not change.
33.1. RS232, RS485 connector (plug in side view) 33.2. Ethernet connector (plug in side view) 33.3. USB connector (plug in side view) 33.4.
34. SPECIFICATION OF ADDITIONAL MODULES Apart from standard interface, it is possible to equip terminals PUE 5 with additional module increasing functionality of devices: • • • • • Weighing module MW-02 8 inputs / 8 outputs module WE 8, 4 inputs / 4 outputs module WE 4, Analogue output module of AN series, Interface Profibus DP V1. 34.1. Weighing module MW-02 The weighing module MW-02 is responsible for the whole process of weighing.
34.1.1. Module technical specification Maximal number of convertor units 8 388 608 OIML class III Number of verification units 6000e Maximal signal gain 19mV Maximal voltage per verification unit 3,3 µV Minimal voltage per verification unit 1µV Minimal strain gauge impedance 90Ω Maximal strain gauge impedance 1200Ω Strain gauge converter supply voltage 5V Strain gauge converters connection 4 or 6 wires + shield 34.1.2.
34.1.3.
SIGNAL FROM WEIGHING MODULE SIGNAL FROM TENSOMETER E EKRAN REF+ REFIN+ IN+5V AGND OUTPUT+ OUTPUTINPUT+ INPUT- NOTICE (according to the rule of connecting shields) (JP1 solder) (JP2 solder) • Connection of strain gauge cable shield With connection between the shield and the load cell body (manufacturer feature) Without connection between the shield and the load cell body (manufacturer feature) The scale with housing or stainless connected with the platform via cable.
34.1.4. The way of installing inside PUE 5 Module is designated for assembly inside terminal PUE5. Module with address 1 is assembled to main board of terminal to J13(Slot 0), module with address 2 is assembled to connector J11 (Slot 1). For the second module MW-02 gland is installed on casing lid, 3m cable ended with insulated conductors is led through gland. Installing procedure: 1. Unplug the terminal (remove the plug from socket 230V); 2. Unscrew the casing lid (back part of terminal casing); 3.
Pcs. Component name 1 Motherboard 2 Weighing module 3 Round ferrite core 4 Ferrite core 4. While installing the module in the plastic socket on the main board turn you attention to the position of the module (see the figure below), 5. The strain gauge cable led through additional gland PG7 situated on the back wall next to the gland of the main platform cable; 6. Pull a ferrite core on the cable (appropriate internal diameter); 7. Wind wires of the cable on the round ferrite core (5 coils); 8.
34.2.1. Technical specification Parameters of outputs Quantity of outputs 8 Type of outputs Reed operation contacts Wire diameter 0,14 - 0,5mm2 Maximal load-current contact capacity 0,2A DC Maximal forward voltage 50V DC Parameters of inputs Quantity of inputs 8 Input type Optoinsulated Wire diameter 0,14 – 0,5mm2 Control voltage range 5 -24V DC 34.2.2. Installing method in PUE 5 terminal Module is designated for assembly inside terminal PUE5.
34.2.3. I/O diagram WE8 inputs diagram WE8 outputs diagram 34.2.4.
GND IN OUT 1 OUT 2 OUT 3 OUT 4 OUT 5 OUT 6 OUT 7 OUT 8 COMMON 9 10 11 12 13 14 15 16 17 (18) YELLOW GREEN 34.3. WE 4 - 4 inputs / 4 outputs module WE 4 module comprises 4 optoinsulated inputs and 4 optoinsulated outputs of reed relays, does not require installing additional modules on the main board of the terminal. The input / output wires are led out via a gland on the back wall of the housing (3m length). 34.3.1.
34.3.2. Colours of cables for I/O PT0083: LEADS NUMBER (FOR JZ-500 108G0,5) SIGNAL IN 1 IN 2 IN 3 IN 4 GND IN OUT 1 OUT 2 OUT 3 OUT 4 COMMON 1 2 3 4 5 6 7 8 9 (10) YELLOW GREEN 34.3.3. Installing method in PUE 5 terminal WE 4 modules are equipped in one cable, for inputs and outputs. Installing procedure: 1. 2. 3. 4.
34.4. Analogue output module AN series Module of analogue outputs Module accessible in three configurations: • • • Voltage output AN 0-10V Current output AN 4-20mA Current output AN 0-20mA 34.4.1. Technical specification Work modes Resolution Current output resistance Voltage output resistance Power supply 4 - 20mA , 0 - 20mA, 0 - 10V 16 bit <500 >400 24V DC (12 - 30V DC) max 40mA 34.4.2. The way of installing inside PUE 5 These modules are intended to mount inside PUE 5.
4. During installation turn your attention to plastic columns. They should be placed one side in mounting holes in the main board and the other side in the mounted module; 5. Led the PT0015 cable through one of the free glands; 6. Connect the PT0015 cable to J3 on the analogue module according to the description below; 7. Connect the PT0015 cable shield to the housing (screwed terminator, 4mm diameter); 8.
34.4.4.
34.5. Profibus interface DP V1 Profibus Plug-In Module PUE5 weighing terminals can be optionally equipped in Profibus Plug-In Modules of DPV1 standard with connectivity via the uniform AnybusCompactCom. A detailed description of the interface is in a separate manual Profibus PUE5. 34.5.1.
Installing procedure: 1. Unplug the terminal from mains; 2. Unscrew and take off the back wall of the housing; 3. Install your module in J9 on the main board; 4. Screw down the interface using 2 screws; 5. Set the DIP-SWITCH DP1 (RS485 – CF MODULES) in position OFF, Switch settings DP1: 1 – ON 6. Set SW1 to 115,2 kbs.
Switch settings SW1: 1 – ON 2 – ON 3 – OFF 7. Connect socket SUB-D with a cable, 8. Remove the hole plug from one of the free glands and led the cable through; 9. The cable connect to the group of wires (unhook band clips fastening the group of wires, lay the cable and hook the band clips). Band clips of multiple usage; 10. Screw down the back wall. 11. Set in the MS Windows Device manager, USB serial port on COM4 35.
36.
37. APPENDIX A – SETTING A BARCODE SCANNER 1. RADWAG scales use RS232 interface with transmission in one direction to communicate with a barcode scanner. It requires only two wire connection. So barcode scanners needs to be equipped with RS232 with disabled hardware and software flow control. 2. Both the terminal and a scanner can have transmission parameters set. Both devices should have set the same parameters: baud rate, data bits, parity control, number of stop bits. e.g.
PRODUCENT WAG ELEKTRONICZNYCH RADWAG WAGI ELEKTRONICZNE 26 – 600 Radom, Bracka Street 28 Main line tel. +48 48 38 48 800, tel./fax. + 48 48 385 00 10 Selling department + 48 48 366 80 06 www.radwag.