Owner manual
Possible answers: 
PC_A_”Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,PC,K1,K0,DH,UH, 
ODH,OUH,S1,S0,OT,UT" – command executed, the indicator 
have sent all the implemented commands. 
28.5. Manual printouts / automatic printouts 
Users can general manual or automatic printouts from the scale. 
•  Manual printouts can be performed after loading the pan and 
stabilizing indication by pressing 
. 
•  Automatic printouts can be performed only after loading the 
pan and stabilizing indication. 
Notice: 
If a scale is verified printouts of immediate values are blocked. 
Format frame: 
1  2  3  4 -12  13  14 15 16  17  18 
stability space sign mass space  unit  CR  LF 
Stability character 
[space] if stable 
[?] if not stable 
[^] if an indication over the range 
[v] if fan indication below the range 
sign 
[space] for positive values or 
[-] for negative values 
mass 
9 characters justified to the right 
unit 
3 characters justified to the left 
command 
3 characters justified to the left 
Example 1: 
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – the printout generated from 
the scale after pressing ENTER/PRINT. 
- 148 -