Owner manual
- 145 -
28.4.12. Switch on continuous transmission in current unit 
Syntax: CU1 CR LF 
Possible answers: 
CU1_I CR LF 
- command comprehended but cannot be executed 
CU1_A CR LF 
MASS FRAME 
- command comprehended and in progress
- mass value in current unit is returned 
Frame format: 
1 2 3  4  5  6  7-15  16  17 18 19 20 21 
S U  I  stability  space sign mass space  unit  CR LF 
28.4.13.  Switch off continuous transmission in current unit 
Syntax: CU0 CR LF 
Possible answers: 
CU0_I CR LF 
- command comprehended but cannot be executed 
CU0_A CR LF 
- command comprehended and executed
28.4.14.  Lock the scale keyboard 
Syntax:  K1 CR LF 
Possible answers: 
K1_I CR LF 
- command comprehended but cannot be executed 
K1_OK CR LF 
- command executed
Caution: 
This command is not remembered after restart 