User guide
EXI-01-04-06-14-GB
- 5 -
PUE C/31H/EX technical data:
PUE C/31H/EX
Housing type stainless steel
Ingress Protection Rating IP66/67
Display type LCD
Keyboard Microswitch (500 000 cycles)
OIML class III
Maximum quantity of A/D converter divisions 8388608
Number of verification intervals 6 000
Maximum input voltage 19,5 mV
Maximum voltage per verification interval
3,25 µV
Minimum voltage per verification interval
1 µV
Ambient temperature -10
o
C ≤ ta ≤ 40
o
C
Minimum load cell impedance
125 Ω
Maximum load cell impedance
1200 Ω
Connection of load cell 4 or 6 cables + shield
Number of ranges 1 or 2 ranges
Power supply - DC IN connector pin 1(4) pin 2(3)
Ui=9V
Ii=0,24A
Pi=0,6W
Ci≈0
Li≈0
I/O marking +5V, E, AGND, REF+, REF-, +IN, -IN
Parameters identical for each clamp
Uo=7,14V
Io=0,24A
Po=0,52W
Li≈0
Ci=6uF
Lo=40uH
Co=10nF
DESIGNATION
II 2G Ex ib IIC T4
KDB 06ATEX250