User manual
99 
Options: 
- 
Automatic correction of reference mass disabled 
-  Automatic correction of reference mass enabled 
Function <  Automatic correction of reference mass> in mode 
<
 Parts counting > is enabled at the moment of estimating the sample 
quantity and signalled by displaying <PCS> and <SMP> (single part/piece 
mass) on the top part of the display. 
There are four criteris of working “Automatic correction of reference 
mass”
 function: 
1.  equilibrium should be reached, 
2.  quantity of parts/pieces should be increased, 
3.  added quantity of parts/pieces should not be greater than double  
number of parts/pieces on the pan, 
4.  the new sample can be different from the old sample by ± 0.3  
of pcs (absolute value), 
If a user recognises that the sample quantity is adequate the unit mass  
(single piece mass) can be saved (see ch. 26.6 of this manual)  
and disable the function by pressing 
. 
Notice: 
While the parts counting function is active the functionality of 
 button 
is changed. Pressing 
 does not result in printing and saving 
weighings. 
26.2.2. Minimum reference mass 
It concerns the <  Parts counting > working mode 
Users before beginning the procedure of single piese mass evaluation can 
declare “minimum reference mass” i.e. minimum total weight of all pieces 
put on the scale pan expressed in reading divisions. 
Procedure: 
•  Enter parameter group <
 Working modes> according to ch.  
24 of this manual, 