User manual
40 
Procedure: 
•  Enter <  Weighing> according to ch. 14 of the manual, select 
<
 Median Filter> and then set an appropriate value. 
Accessible settings: 
None  - median filter is off 
0.5, 1, 1.5, 2, 2.5  - filtering level to choose 
14.2. Filter 
This filter is intended to suppress continuous mechanical vibrations at the  
cost of stabilization time. 
Procedure: 
•  Enter <  Weighing> according to ch. 14 of the manual, select  
<
 Filter> and then set an appropriate value. 
Accessible settings: 
None, V.Fast, Fast, Average, Slow. 
Notice: 
The higher filtering level the longer stabilization time. 
14.3. Autozero 
The autozero function has been implemented in order to assure precise 
indications. This function controls and corrects „0” indication. 
While the function is active it compares the results continuously with  
constant frequency. If two sequentional results differ less than the declared 
value of autozero range, so the scale will be automatically zeroed and the 
pictograms 
 and  will be displayed. 
If AUTOZERO is disabled zero is not corrected automatically. However,  
in particular cases, this function can disrupt the measurement process e.g.  
slow pouring of liquid or powder on the weighing pan. In this case, it is 
advisable to disable the autozero function.  