User manual
190 
34.4.20.  Send all implemented commands 
Syntax:  PC CR LF 
Possible answers:  
PC A "Z,T,S,SI,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,SIA,SS, 
PC" – command executed, the indicator have sent all the implemented  
commands. 
34.5. Manual printouts / automatic printouts 
Users can general manual or automatic printouts from the scale.  
•  Manual printouts can be performed after loading the pan  
and stabilizing indication by pressing 
.  
•  Automatic printouts can be performed only after loading  
the pan and stabilizing indication. 
Format frame: 
1  2  3  4 -12  13  14  15  16  17  18 
stability  space  sign  mass  space  unit  CR  LF 
Stability character 
[space] if stable 
[?]  if not stable 
[^] if an indication over the range 
[v]  if fan indication below the range 
sign 
[space] for positive values or  
[-]   for negative values 
mass 
9 characters justified to the right 
unit 
3 characters justified to the left 
command 
3 characters justified to the left 
Example: 
_ _ _ _ _ _ 1 8 3 2 . 0 _ g _ _ CR LF – the printout generated from  
the scale after pressing 
. 