User manual
188 
34.4.13.  Switch off continuous transmission in current unit 
Syntax: CU0 CR LF 
Possible answers: 
CU0_I CR LF 
- command comprehended but cannot be executed 
CU0_A CR LF 
- command comprehended and executed 
34.4.14.  Set lower threshold 
Syntax:  DH_XXXXX CR LF, where: XXXXX – mass format 
Possible answers: 
DH_OK CR LF 
- command executed 
ES CR LF 
- command not comprehended (wrong mass format) 
34.4.15.  Set upper threshold 
Syntax:  UH_XXXXX CR LF, where: XXXXX – mass format 
Possible answers: 
UH_OK CR LF 
- command executed 
ES CR LF 
- command not comprehended (wrong mass format) 
34.4.16.  Read lower threshold 
Syntax:  ODH CR LF 
Possible answers:  DH_MASA CR LF - command executed 
Frame format: 
1  2  3  4-12  13  14  15  16  17  18  19 
D  H  space  mass  space  unit  space  CR  LF 
Mass  - 9 characters justified to the right 
Unit   - 3 characters justified to the left 