34.4.10.  Switch on continuous transmission in basic unit 
Syntax: C1 CR LF 
Possible answers: 
C1_I CR LF 
- command comprehended but cannot be executed 
C1_A CR LF 
MASS FRAME 
- command comprehended and in progress 
- mass value in basic unit is returned 
Frame format: 
1  2  3  4  5  6  7-15  16  17  18  19  20  21 
S  I  space  stability  space  sign  mass  space  unit  CR  LF 
34.4.11.  Switch off continuous transmission in basic unit 
Syntax: C0 CR LF 
Possible answers answers: 
C0_I CR LF 
- command comprehended but cannot be executed 
C0_A CR LF 
- command comprehended and executed 
34.4.12.  Switch on continuous transmission in current unit 
Syntax: CU1 CR LF 
Possible answers: 
CU1_I CR LF 
- command comprehended but cannot be executed 
CU1_A CR LF 
MASS FRAME 
- command comprehended and in progress 
- mass value in current unit is returned 
Frame format: 
1  2  3  4  5  6  7-15  16  17  18  19  20  21 
S  U  I  stability  space  sign  mass  space  unit  CR  LF 