User manual
182 
Notice: 
1.  Each command have to be terminated in CR LF; 
2.  The best Policy for communication is not sending another  
command until the former answer has been received.  
34.3. Respond message format 
After sending a request message you can receive: 
XX_A CR LF 
command accepted and in progress 
XX_D CR LF  
command completed (appears only after XX_A) 
XX_I CR LF  
command comprehended but cannot be executed 
XX _ ^ CR LF 
command comprehended but time overflow error appeared 
XX _ v CR LF 
command comprehended but the indication below the 
XX _ OK CR LF 
Command done 
ES_CR LF 
Command not comprehended 
XX _ E CR LF 
error while executing command – time limit for stable result 
exceeded (limit time is a descriptive parameter of the scale) 
XX  - command name 
_    - substitutes spaces 
34.4. Command’s description 
34.4.1. Zeroing 
Syntax Z CR LF 
Possible answers: 
Z_A CR LF 
Z_D CR LF 
- command accepted and in progress 
-
 command completed 
Z_A CR LF 
Z_^ CR LF 
- command accepted and in progress 
-
 command comprehended but zero range overflow appeared 
Z_A CR LF 
Z_E CR LF 
- command accepted and in progress 
-
 time limit for stable result exceeded 
Z_I CR LF 
- command comprehended but cannot be executed 