User manual
157 
Process course: 
•  Press <  Local parameters> key to enter the setting of local 
parameters in density mode. 
•  Press <  Standard liquid> key to select the auxiliary liquid used  
for the density determination procedure. If the auxiliary liquid is other 
than “Water” or “Alcohol”, then select “Other” from the available list. 
•  Press < Temperature> i key and specify the temperature of the 
auxiliary liquid [°C]. Accept inserted value by pressing 
 key. 
•  If „Other” liquid has been selected, then press <  Standard liquid 
density> insert the density value of the auxiliary liquid expressed in 
[g/cm
3
] for the specified temperature of the measurement. 
•  On specifying the parameters exit to main menu of the density  
mode by pressing 
 key. 
•  In order to start density determination of porous body, press <  
Determine porous body density> key, using a pre-defined hot keys 
on terminal’s display. 
Caution: 
If in the local parameters, <
 Ask about sample number> function  
is enabled, then on initiating the process, the system asks for inserting  
the number of tested sample. The inserted sample number is linked to  
the process data and saved in the database. 
•  The first phase of the density determination of a porous body in the 
air. Place the object on the scale’s weighing platform. On obtaining 
stable measurement result, accept it by pressing 
 key, 
•  Then immerse the tested porous body in oil to fill its pores. Place  
the tested porous body impregnated with oil on scale’s weighing 
platform. On obtaining stable measurement result, carry out the 
measuring process, and accept it by pressing 
 key, 
•  In the third step, weight the tested porous object impregnated with  
oil in the auxiliary liquid. Place the tested object in the auxiliary liquid, 
on scale’s weighing platform and immerse both in the auxiliaty liquid.   
On obtaining stable measurement result, carry out the measuring 
process, and accept it by pressing 
 key, 