User manual
152 
Process course: 
•  Press <  Local parameters> key to enter the setting of local 
parameters in density mode. 
•  Press <  Sinker volume> in order to insert the volume of the 
sinker (given in [cm
3
]) which is immersed in the tested liquid. 
•  On specifying the parameters exit to main menu of the density  
mode by pressing 
 key. 
•  In order to start density determination of liquid, press  
<
 Determine liquid density> key, using a pre-defined hot  
keys on terminal’s display. 
Caution: 
If in the local parameters, <
 Ask about sample number> function  
is enabled, then on initiating the process, the system asks for inserting  
the number of tested sample. The inserted sample number is linked to  
the process data and saved in the database.   
•  The first phase of the density determination of liquid is measurement 
of sinker’s mass in the air. Place the sinker on the scale’s weighing 
platform. On obtaining stable measurement result, accept it by 
pressing 
 key. 
•  Next, place on scale’s weighing platform tested liquid with immersed 
sinker. On obtaining stable measurement result, carry out the 
measuring process, and accept it by pressing 
 key. 
•  On carrying out the second measurement, the scale determines 
density of tested liquid and indicates the result on terminal’s display.  
If the user needs a report from the density determination process, it 
can be printed on a plugged printer by pressing 
 key. The 
density determination process is then completed.  
•  A report from density determination of liquid is saved in <  
Density> database. The report is named by process date.  