User manual
116 
Platform 2: 
Icon  Step  Value  Description 
1. [CF] 
Flags 
condition 
Signal flag 1 – „1” 
Checking condition, whether signal flag 1 is set to 
value “1” – i.e. checking whether desired part of the 
process has been already realized on the weighing 
platform no. 1. If Yes, then dosing process on the 
weighing platform no. 2 shall start. 
2. [T] 
Tare 
Tare 
Tarring the weighing platform no. 2 
3. [TI] 
Delay 
[5s] Open valve 
Water 
Pending for opening main valve of the container with 
“Water” 
4. [DA] 
Automatic 
batching 
2 kg [Water] 
Automatic dosing of ingredient “Water” to obtain 2 kg 
mass in single-step dosing mode – Output 6 controls 
the dosing valve (set for Water in Products database) 
6. [TI] 
Delay 
[5s] Close valve 
Water 
Pending for closing the main valve of the container 
“Water” 
7. [O] 
Outputs 
Output 12 – „1” 
Output 12 akes high status (“1”) – signal confirming 
carrying out the dosing process on the weighing 
platform no. 2 is lit up 
8. [F] Set 
flags 
Signal flag 1 – „1” 
Signal flag 2 – „1” 
Setting the characteristic point of the process, which is 
a condition for carrying out part of the process on the 
weighing platform no. 1 
8. [TI] 
Delay 
[10s] Dosing 
completed 
Displaying a text message on the terminal confirming 
completion of dosing process 
An instance of mixed dosing (manual and automatic) has been extended by 
signalling its corresponding phases to present the possibilities of the working 
mode <Dosing>. 
28.7. Reporting of completed dosing processes 
On completing each dosing process, the terminal automatically generates  
a report on that process.  
Notice: 
Submenu: “
 Devices /   Printer /  Printouts /  Dosing 
report printout template” enables optional modifying of the report  
template (see ch. 16.2.3 of this user manual). 