User manual
100
• Choose „ Parts counting / Minimum reference mass”
and then set an appropriate value.
Accessible settings: 1 d, 2 d, 5 d, 10 d.
Notice:
While the procedure of evaluation the mass of single piece the the mass
of all pieces put on the pan is lower than the value declared in parameter
„
Minimum reference mass”, the following warning message will
be displayed: < Too low sample mass>.
26.3. Setting a reference unit by entering known piece mass
Procedure:
• Enter mode <
Parts counting> according to ch. 26.1
of this manual,
• Press
(enter piece mass), then an editing field is displayed
<Reference Unit> with the screen keyboard,
• Enter a value and confirm it by pressing
, which causes starting
<
Parts counting > with automatic setting the reference unit.
Notice:
1. In case of entering a reference unit higher then the maximum weighing
range of the main scale the program will display a message box:
<Value too high>,
2. In case of entering the single piece mass lower than 0.1 d, the
program will display a message box: <Value to small>.
26.4. Setting a reference unit by weighing a sample
Procedure:
• Enter mode < Parts counting> according to ch. 26.1
of this manual,
• If pieces are weighed in a container it needs to be put on the pan
and tarred,