Choose Parts counting / Minimum reference mass”
and then set an appropriate value.
Accessible settings: 1 d, 2 d, 5 d, 10 d.
Notice:
While the procedure of evaluation the mass of single piece the the mass
of all pieces put on the pan is lower than the value declared in parameter
Minimum reference mass, the following warning message will
be displayed: < Too low sample mass>.
26.3. Setting a reference unit by entering known piece mass
Procedure:
Enter mode <
Parts counting> according to ch. 26.1
of this manual,
Press
(enter piece mass), then an editing field is displayed
<Reference Unit> with the screen keyboard,
Enter a value and confirm it by pressing
, which causes starting
<
Parts counting > with automatic setting the reference unit.
Notice:
1. In case of entering a reference unit higher then the maximum weighing
range of the main scale the program will display a message box:
<Value too high>,
2. In case of entering the single piece mass lower than 0.1 d, the
program will display a message box: <Value to small>.
26.4. Setting a reference unit by weighing a sample
Procedure:
Enter mode < Parts counting> according to ch. 26.1
of this manual,
If pieces are weighed in a container it needs to be put on the pan
and tarred,