User manual

- 40 -
P1.5 CALt - Time of automatic adjustment (balances with internal
adjustment system)
Determination of time interval, after which automatic adjustment process is
initiated.
P1.6 CALr printout of adjustment report
Printout activated on completion of automatic adjustment procedure.
balances with internal adjustment balances with external adjustment
no report printout disabled
YES report printout enabled
Go back to weighing mode with procedure of saving changes
(see point - 8. 2. return to weighing).