User manual

- 25 -
YES - mode enabled; no - mode disabled
On selecting, press PRINT key. The balance returns to displaying
submenu name.
Go back to weighing mode with procedure of saving changes
(see point - 8. 2. return to weighing).
9.4. Dual range balance (APP 6/35/2C)
Balance APP 6/35/2C series is a dual range weighing instrument. The accuracy of
the I weighing range is d
1
=1g, and of the II weighing range it is d
2
=5g.
Transfer from weighing with the accuracy of the I weighing range to weighing with
the accuracy of the II weighing range takes place automatically on exceeding Max
1
point at 6kg (with no operator activity needed). On switching to weighing with the
accuracy of the II weighing range, the display signals it by indicating II pictogram on
its right side.
From now on the balance weighs mass with the accuracy of the II weighing range.
In order to return to weighing in the accuracy of the I weighing range:
Take the weighed load off the weighing pan
As the indication returns to zero and pictograms
0
and are lit
press
0/T
key on balance’s keyboard.