User manual
- 208 -
User log out
Format: LOGOUT CR LF
Accessible responses:
LOGOUT OK CR LF
- command understood, user is logged out
ES CR LF
- command not recognized (error in format)
Set balance mode
Format: PROFILE_Name CR LF where: _ - space (Name must be accordant
with the name form entered in a balance – lowercase and uppercase letters)
Accessible responses:
PROFILE OK CR LF
- command understood, new PROFILE activated
LOGIN ERRROR CR LF
- command understood, an error in PROFILE name,
log in failed
ES CR LF
- command not recognized (error in format)
Send all implemented commands
Format: PC CR LF
Accessible responses:
PC_A_”Z,T,S,SI,SIA,SU,SUI,C1,C0,CU1,CU0,DH,ODH,UH,OUH,OT,UT,
LOGIN,LOGOUT,PC" – command carried out, terminal has sent all
implemented commands.
36.3. Manual printout / Automatic printout
Balance 3Y series enables generating manual or automatic printouts.
• Manual printout: press key on stabilization of indication
(measurement result),
• Automatic printout is generated automatically, in accordance to the settings
of automatic printout (see point 10.8 of this user manual).
The content of a printout depends on setting in menu <Standard printout> -
<Weighing printout template>(see point 15.5 of this user manual).