User manual

- 207 -
Set mass value of a single (only for PARTS COUNTING)
Format: SM_XXXXX CR LF, where: _ - space, XXXXXmass format
Accessible responses:
SM_OK CR LF
SM_I CR LF
- command carried out
-
command understood but not accessible at this
moment
(e.g.: other working mode, not PARTS
COUNTING
)
ES CR LF
- command not recognized (incorrect mass format)
Set target mass value (e.g. for DOSING)
Format: TV_XXXXX CR LF, where: _ - space, XXXXXmass format
Accessible responses:
TV_OK CR LF
TV_I CR LF
- command carried out
-
command understood but not accessible at this
moment
(e.g.: WEIGHING mode)
ES CR LF
- command not recognized (incorrect mass format)
Set reference mass value (e.g. for PERCENT SETUP)
Format: RM_XXXXX CR LF, where: _ - space, XXXXXmass format
Accessible responses:
RM_OK CR LF
RM_I CR LF
- command carried out
-
command understood but not accessible at this
moment
(e.g.: other working mode, not PERCENT
SETUP
)
ES CR LF
- command not recognized (incorrect mass format)
User log in
Format: LOGIN Name, Password CR LF (Name and password have to be
inserted as they are set in the balance small and capital letters)
Accessible responses:
LOGIN OK CR LF
- command understood, new user is logged in
LOGIN ERRROR CR LF
- command understood, an error in name or
password occurred, log in failed
ES CR LF
- command not recognized (error in format)