User's Manual
Table Of Contents
- 1. General Information
- 2. Transport and Storage
- 3. Unpacking and Installation
- 4. Start-Up
- 5. Keyboard – Buttons Functions
- 6. Weighing Mode Home Screen
- 7. Operating Balance Menu
- 8. Weighing Operation
- 8. Adjustment
- 9. Working Modes – General Information
- 10. Databases
- 11. Communication
- 12. Peripheral Devices
- 13. Communication Protocol
- 14. Error Messages
- 20 -
8. WEIGHING OPERATION
Load a weighed object on a balance weighing pan. On stabilization of weighing result, indicated by
stability marker
visible on the left side of balance display, read the measurement result.
Record / printout of the measurement result is available on pressing <PRINT> key:
• for verified balances – only stable measurement result can be saved or printed (stability marker
visible on balance’s display),
• for non-verified balances – stable or unstable measurement result can be saved or printed
(regardless of stability marker
absence). If unstable measurement result is printed then it is
accompanied by question mark <?> in front of printed mass value.
8.1. GOOD WEIGHING PRACTICE
In order to ensure long lasting use of a balance plus correct and reliable measurement of weighed
loads, follow below procedures:
• Start the balance with no load on the weighing pan (permissible value of load on the weighing
pan on balance start is ±10% of its maximum capacity).
• Load the weighing pan steadily avoiding shocks:
YES
NO
• Place weighed loads centrally on the weighing pan:
YES
NO
• Avoid side loading, in particular side shocks:
NO
NO