User manual

The Hosts list is a list of destinations that the Radlink system could forward
the DICOM image to. To instruct the Radlink system to forward the image to
a host destination, check the check box for that destination.
The above figure shows a modality that has been configured to send to the
destinations Radlink and LOCAL PACS. One could also send the image to
the Dr. Jim Bob by selecting it as well in the above example.
The port setting is the port that the Radlink system will use to receive the
DICOM images from that modality. The modality should be configured to
send to the IP that is defined on this machine, the machine housing the
modality router.
The AE title at the sending modality must be set to the Router AE Title. The
port setting in the example above is 1005. There cannot be duplicate
modality AE Titles and the modality ports cannot conflict with other port
settings within the machine. The range of ports that can be assigned are
from 1 to approximately 65,000.
To set up other modalities, follow the above procedure for adding and
configuring a modality. One could configure two different modalities to send
to the same Radlink modality port by configuring those modalities to send to
the same Radlink port.
Note that if the Radlink DICOM Router is shut down, the Radlink system will
not be able to receive images. The Radlink DICOM Router is started
automatically upon machine startup.
Figure 3-49 – Radlink Router Destination Status Screen
62