The CR Pro is fused for 5 amp domestic and 3 amp European. Fuses are
located in the AC power input fixture on the back, bottom, right hand side of
the CR Pro.
FUSE – The input power is fused at the AC power input connector at the
bottom, right rear of the CR Pro cabinet. The fuse(s) is located in a box directly
above the input connector and is a integral part of the connector assembly. To
remove the fuse box, perform the following:
1. Turn off power to the device.
2. Remove the AC line cord from the connector.
3. Using a small flat tip device, insert the tip under the tab at the bottom
of the box above the connector and gently pry the box open.
4. Inside the box are two (2) fuses as follows:
Domestic 100 – 120VAC 5A Fuse(s) (standard)
European 220 – 240VAC 3A Fuse(s) (standard)
Replace the fuse(s) with the value required for your model CR Pro
scanner and replace the box in its receptacle making sure the box is fully
inserted.
Replace the AC line code and turn the power switch on. If the fuse blows
again, turn the power switch OFF and immediately call for service.
● ENVIRONMENT –
Temperature – 60ºF (15ºC) to 80ºF (27ºC) Operating
40ºF (4ºC) to 90ºF (33ºC) Non-operating
Relative Humidity – 20% to 80% non-condensing
Light – Phosphor plates are light sensitive and are erased if exposed to light. The
CR Pro cassettes are light protected. However, light can penetrate the CR Pro
light seal if bright light or sunlight is directed to the cassette slot on top of the CR
Pro. Ambient light in the proximity of the CR Pro should be kept to a minimum.