User manual

40” - single-phase rectified generator, increase the 8 mAs by 1.5 * 8 = 12
mAs
40” - single-phase unrectified generator, increase the 8 mAs by 2.0 * 8
=16 mAs
The calibration procedure just requires enough x-rays to create a uniform
"flood field" exposure on the plate -- there is no "right" or "wrong"
technique -- just guidelines for creating a uniform field.
1. Start Application – The Radlink application should be running, if not,
select the Radlink application from the list of programs and start it
on the CR Pro
2. Load Plate Scan – Insert the exposed cassette in the CR Pro machine.
3. Start Calibration - Select the calibration selection under the TOOLS
menu item.
4. Calibration - The CR Pro will scan the exposed plate in order to calibrate
the intensity to allow for the best possible images to be acquired from the
plate. Once scanned, the system will then go through a second pass and
erase the plate to normal levels.
1. Finished – The CR Pro is now calibrated and you can begin
using it for processing X-ray
2. Main Screen